Published date: 7 June 2019
Closed opportunity - This means that the contract is currently closed. The buying department may be considering suppliers that have already applied, or no suitable offers were made.
Contract summary
Industry
Laboratory, optical and precision equipments (excl. glasses) - 38000000
Location of contract
TW11 0LW
Value of contract
£0
Procurement reference
tender_209740/774748
Published date
7 June 2019
Closing date
9 July 2019
Contract start date
1 August 2019
Contract end date
31 December 2019
Contract type
Supply contract
Procedure type
Open procedure (above threshold)
Any interested supplier may submit a tender in response to an opportunity notice.
This procedure can be used for procurements above the relevant contract value threshold.
Contract is suitable for SMEs?
Yes
Contract is suitable for VCSEs?
No
Description
The scope of this procurement is for a complete laser interferometer system (commercially referred to as a Vibrometer). This will include a laser interferometer instrument (laser head and driving boards), necessary third-party electrical equipment required to acquire signal data (oscilloscope, DAQ boards, PC-interface boards), software to operate the instrument on an NPL PC (or otherwise the tenderer to supply a computer where necessary) and physical hardware in order to integrate the instrument into the existing facilities at NPL (stand, tripod, objective lenses).
Background to the requirement:
The NPL requirement is for a laser interferometer for the displacement measurement of a water-borne reflective membrane incident upon by ultrasonic tone bursts. The purpose is to provide accurate and reliable picometre displacement measurements of the membrane, when driven at frequencies between 0.1 and 100 MHz, for the purposes of calibrating ultrasonic hydrophones.
The measurement of displacement is expected to be robust to environmental laboratory noise including acoustic amplifier noise, typical RF bursts from pulsed lasers, air movement through air conditioning, airborne sound and environmental vibration. Access should be available to the raw data to allow for independent post-processing.
The system should not be so bulky or sensitive that it cannot easily be moved once installed. The system will be expected to be used in multiple laboratories and as such must be suitable for efficient transport and set up.
More information
Links
-
- https://lupc.bravosolution.co.uk/web/login.shtml
- Tender notice
- Please ensure that you register on Bravo Solutions (link below) to obtain the documents and submit your interest / intention to bid.
Additional text
-
Please do not contact the person listed on this notice. Documents can only be obtained through Bravo Solutions.
Please ensure that you register on Bravo Solutions (link below) to obtain the documents and submit your interest / intention to bid.
Share this notice
Closing: 9 July 2019
All content is available under the